Differential photoacoustic spectroscopy for flow gas detection based on single microphone

Lujun Fu,Jiangshan Zhang, Yufeng Pan,Ping Lu

Photoacoustics(2024)

引用 0|浏览0
暂无评分
摘要
Differential photoacoustic spectroscopy (PAS) for flow gas detection based on single microphone is innovatively proposed and experimentally demonstrated. Unlike the traditional systems, only one microphone is used to suppress flowing gas noise. Wavelength modulation spectroscopy and second harmonic detection technique are applied in this PAS system with Q-point demodulation for acetylene (C2H2) gas detection. The experiment is conducted at 1atm and 300K. Different concentrations and flow rates of C2H2 from 0 sccm to 225 sccm are detected by using nitrogen (N2) as the carrier gas, which indicates that the system can respond well to flowing gases while maintaining the noise at the same level. The system response time decreases to 3.58s while the gas velocity is 225 sccm. The detection limit of 43.97 ppb with 1s integration time and normalized noise equivalent absorption (NNEA) coefficient of 4.0 × 10-9cm-1·W·Hz-1/2 is achieved at the flow rate of 225 sccm. The firstly proposed differential PAS based on single microphone greatly simplifies the system structure for flow gas detection, which provides a novel route for development of PAS with significant practical implementation prospects.
更多
查看译文
关键词
Flow gas detection,Online detection,Differential spectroscopy,Photoacoustic spectroscopy,Single microphone,C2H2