Transcatheter Closure of Perimembranous Ventricular Septal Defect Using KONAR-MF™: A Multicenter Experience

Pediatric Cardiology(2024)

引用 0|浏览0
暂无评分
摘要
Transcatheter closure of perimembranous ventricular septal defect (PmVSD) is an established procedure. However, the occurrence of complete heart block limits its scope. The newer KONAR-MF™ occluder has specific design characteristics that may improve the safety of PmVSD closure. The objective of the study was to describe the efficacy and mid-term follow-up of transcatheter closure of PmVSD using KONAR-MF™. The study was conducted prospectively in 3 Indian centers (January 2018–December 2022). PmVSD closure was done by both antegrade and retrograde methods, and patients were followed up at 1, 3, 6, 12 months, and annually after that. 121 out of 123 patients were included with the following characteristics: median age 4.4 (0.18–40) years; weight 15 (2.1–88) kg; mean Qp/Qs ratio 1.87 ± 0.52 and pulmonary artery mean pressure: 22 ± 6.9 mmHg. The procedure was successful in all but 3; the device was removed due to significant residual shunt (n = 2) and new development of aortic regurgitation (AR) (≥ mild) in 1. The median defect size was 5.2 (2.5–12) mm. Device sizes from 6/4 to 14/12 were deployed (median fluoroscopy time 13.3 min; range 3.6–47.8). Shunt occlusion rates were 90
更多
查看译文
关键词
Perimembranous ventricular septal defect,Transcatheter device closure,Occluder,KONAR-MF,Complete heart block