Approach for contact medical device development via integrated testing, skeletal muscle modeling, and finite element analysis.

Conor Shanley,Q Jane Wang, Bruce Livingston

Journal of the mechanical behavior of biomedical materials(2024)

引用 0|浏览0
暂无评分
摘要
Development of novel medical devices for the treatment of musculoskeletal pain associated with neuro-muscular trigger points requires a model for relating the mechanical responses of in vivo biological tissues to applied palliative physical pressures and a method to design treatments for optimal effects. It is reasonable to hypothesize that the efficacy of therapeutic treatment is proportional to the maximum tensile strain at trigger point locations. This work presents modeling of the mechanical behavior of biological tissue structures and treatment simulations, supported by indentation experiments and finite element (FE) modeling. The steady-state indentation responses of the tissue structure of the posterior neck were measured with a testing device, and an FE model was constructed using a first-order Ogden hyperelastic material model and calibrated with the experimental data. The error between experimental and FE-generated displacement-load curves was minimized via a two-stage optimization process comprised of an Optimal Latin Hypercube design-of-experiments analysis and a Bayesian optimization loop. The optimized Ogden model had an initial shear modulus (μ) of 5.16 kPa and a deviatoric exponent (α) of 11.90. Another FE model was developed to simulate the deformation of the tissue structures in the posterior neck adjacent to the C3 vertebrae in response to indentation loading, in order to determine the optimal location and angle to apply an indentation force for maximum therapeutic benefit. The optimal location of indentation was determined to be 28° lateral from the sagittal plane along the surface of the skin, measured from the centerline of the spine, at an angle of 8° counterclockwise from the surface normal vector. The optimized spatial orientation of the indentation corresponded to the average of the maximum principal strain across the deep muscle region of the model.
更多
查看译文
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要