Comparison of four in vitro test methods to assess nucleus pulposus replacement device expulsion risk

JOR SPINE(2024)

引用 0|浏览0
暂无评分
摘要
BackgroundNucleus replacement devices (NRDs) are not routinely used in clinic, predominantly due to the risk of device expulsion. Rigorous in vitro testing may enable failure mechanisms to be identified prior to clinical trials; however, current testing standards do not specify a particular expulsion test. Multiple methods have therefore been developed, complicating comparisons between NRD designs. Thus, this study assessed the effectiveness of four previously reported expulsion testing protocols; hula-hoop (Protocol 1), adapted hula-hoop (Protocol 2), eccentric cycling (Protocol 3), and ramp to failure (Protocol 4), applied to two NRDs, one preformed and one in situ curing.MethodsNucleus material was removed from 40 bovine tail intervertebral disks. A NRD was inserted posteriorly into each cavity and the disks were subjected to one of four expulsion protocols.ResultsNRD response was dependent on both the NRD design and the loading protocol. Protocol 1 resulted in higher migration and earlier failure rates compared to Protocol 2 in both NRDs. The preformed NRD was more likely to migrate when protocols incorporated rotation. The NRDs had equal migration (60%) and expulsion (60%) rates when using unilateral bending and ramp testing. Combining the results of multiple tests revealed complimentary information regarding the NRD response.ConclusionsAdapted hula-hoop (Protocol 2) and ramp to failure (Protocol 4), combined with fluoroscopic analysis, revealed complimentary insights regarding migration and failure risk. Therefore, when adopting the surgical approach and animal model used in this study, it is recommended that NRD performance be assessed using both a cyclic and ramp loading protocol. Current nucleus replacement device (NRD) testing standards do not include a specific expulsion test as a requirement. This has resulted in multiple methods being developed inhibiting comparison between designs. This in vitro study compared four expulsion testing protocols reported in previous studies. Results demonstrated that it is beneficial to utilize a range of tests to assess NRDs at the pre-clinical stage so that more accurate assessments of expulsion risk can be made. image
更多
查看译文
关键词
biomechanical testing,expulsion,hydrogel,intervertebral disk,nucleus replacement,spine
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要