Severe ST-segment elevation and AV block during pulsed-field ablation due to vasospastic angina — a novel observation

Journal of Interventional Cardiac Electrophysiology(2024)

引用 0|浏览1
暂无评分
摘要
Catheter ablation of atrial fibrillation using non-thermal electroporation represents a promising ablation modality due to its believed superior safety profile. Still, if electroporation is delivered in proximity to a coronary artery, vasospasms can occur. We report the first case of severe right coronary artery vasospasm resulting in ST-segment elevation and AV block despite a remote distance from the ablation site to the right coronary artery, indicating a different mechanism. In this case, electroporation most likely triggered a previously unknown Prinzmetal vasospastic angina in the patient, resulting in the coronary vasospasm. Thus, meticulous monitoring of ST-segment changes following PFA delivery even from regions remote to coronary arteries is required.
更多
查看译文
关键词
Pulsed-field ablation,Coronary vasospasm,Atrial fibrillation,Catheter ablation,Vasospastic angina,Variant angina