0309 Concordance in Electroencephalography-Headband versus Polysomnography by Demographics and Cognitive Status

Brienne Miner, Yulu Pan, Gawon Cho, Chase Burzynski, Lakshmi Polisetty,Margaret Doyle,Lynne Iannone,Melissa Knauert,Thomas Gill,Henry Yaggi

SLEEP(2024)

引用 0|浏览3
暂无评分
摘要
Abstract Introduction Polysomnography (PSG) can be burdensome and alternative measures are needed to advance sleep research. We previously reported that a novel electroencephalography (EEG)-headband outperformed actigraphy when compared to PSG among older persons with sleep complaints. The current study examined the concordance of the EEG-headband with PSG, and variations in concordance by gender, race, and cognitive status. Methods Sixty community-living adults ≥60 years with self-reported sleep complaints completed a 7-day home-based sleep protocol with a novel EEG-headband (Beacon Biosignals) and one night of home-based polysomnography (NOX-A1 Sleep System). We evaluated the concordance of total sleep time [TST] and wake after sleep onset [WASO] between PSG and the EEG-headband using paired t-tests and two-way mixed effects intraclass correlation coefficients (ICC). We examined variations in concordance by gender, race, and the presence of mild cognitive impairment [MCI] (< 13 on the Montreal Cognitive Assessment short form) using Fisher Z-transformation to assess between-group differences in ICCs. Results Mean age was 74 years, 65% were women, 33.3% were people of color, and 35% had MCI. The EEG-headband showed excellent concordance with PSG when measuring TST (absolute difference -9.14 minutes [95% Confidence Interval -23.17, 4.89]; ICC= 0.83 [0.73, 0.89]) and good concordance when measuring WASO (absolute difference -17.11 [-13.71, -2.50]; ICC= 0.64 [0.47, 0.77]). Concordance in TST was significantly different between men and women (ICCs of 0.91 vs 0.67, respectively; p=0.01), but did not vary by race and cognitive status. Concordance in WASO did not vary by gender, race, or cognitive status. Conclusion Among community-living older adults with sleep disturbances, the EEG-headband showed good to excellent concordance with PSG when measuring TST and WASO. The EEG-headband is an acceptable alternative to PSG, including among people of color or persons with MCI. Reduced concordance in TST among women may have been due to differences in fit and/or contact of the headband with the scalp. Support (if any) NIA (K76AG0749505, R03AG073991, P30AG021342)
更多
查看译文
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要