Quantification of Vitamin A in Edible Oils: Comparison of Portable Device iCheck Chroma3 to High-Performance Liquid Chromatography

Food Analytical Methods(2024)

引用 0|浏览0
暂无评分
摘要
Fortification of edible oil with vitamin A is a widely adopted intervention to minimize the effects of vitamin A deficiency in vulnerable groups and mitigate some of its deleterious consequences. Regulatory monitoring is an important prerequisite to ensure that the fortification program is implemented effectively. Standard laboratory analysis methods for vitamin A in oils to assess adequate addition levels remain expensive and time-consuming. Portable testing devices are relatively less expensive in terms of capital investment and cost per test. However, the reliability of results needs to be assured to ensure acceptability and confidence. This study compared a portable device to high-performance liquid chromatography (HPLC) in terms of quantification of vitamin A in both spiked and commercially fortified oils. Nine oils (soybean, palm, cottonseed, rapeseed, corn, peanut, coconut, sunflower, and rice bran oils) were selected and spiked with retinyl palmitate at six different concentrations, and 112 commercially fortified oils were quantified for their vitamin A content using both methods. A good indicator of intra-day and inter-day repeatability (< 10
更多
查看译文
关键词
Vitamin A,Spiked and fortified oils,Portable device,Repeatability,Linearity,Limits of agreement,HPLC,iCheck Chroma3
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要