Antecubital Burn Resulting From Antenna Coupling

Taylor L. Paskey, Mark Ren, Varun Arvind, Wakenda K. Tyler,Jarrod T. Bogue,Robert J. Strauch

JBJS Case Connector(2024)

引用 0|浏览0
暂无评分
摘要
Case: We present the case of a 54-year-old man who underwent elective hip disarticulation complicated by third-degree burn of the left antecubital fossa requiring skin graft. After careful review, it was determined that “antenna coupling” as a result of electrosurgery was the likely cause. We present an experiment demonstrating this phenomenon. Conclusion: Antenna coupling is a real but rare cause of intraoperative burns not previously described in the orthopaedic literature. Care should be taken to avoid coiling or running bovie or other electrosurgical device cords with other metallic cords or corded devices.
更多
查看译文
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要