Intraoperative Extravascular Ultrasound in the Identification of Flow-Limiting Dissections after Balloon Angioplasty in the Femoropopliteal Segment.

Journal of clinical medicine(2024)

引用 0|浏览2
暂无评分
摘要
Background: The BIO REACT study is designed to investigate the incremental value of Extravascular UltraSound (EVUS) added to conventional angiography, compared to conventional angiography only for the identification of Flow-Limiting Dissections (FLD) and to evaluate the safety and efficacy of the REsponse Adapted Combination Therapy (REACT) for the treatment of femoropopliteal lesions. Methods: The primary endpoints were the specificity and sensitivity of EVUS added to angiography for the detection of FLD. Secondary endpoints were primary patency of the REACT therapy within 12 months, fCD-TLR, freedom from MAE, major target limb amputations (mTLA) and survival rates within 24 months. Results: A total of 150 patients were included. EVUS added to angiography had an overall sensitivity of 29% and specificity of 93% for the detection of FLD. There was no PSVR cut-off offering a clinically acceptable trade-off between meaningful sensitivity and specificity values for the detection of FLD. At 12 months, treatment with the REACT resulted in primary patency and fCD-TLR of 81.6% and 94.3%, respectively. In addition, freedom from MAE was 94.3% at 12 months. At 24 months, the survival rate was 94.0%. No mTLA was reported up to the 24-month follow-up. Conclusions: The addition of DUS to angiography showed limited value for detecting FLD in femoropopliteal artery disease.
更多
查看译文
关键词
extravascular ultrasound,flow-limiting dissection,drug-coated balloon,femoropopliteal artery
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要