The Frank Stinchfield Award: Assembly and Dissociation Forces Differ Between Commonly Used Dual Mobility Implants: A Biomechanical Study

Katherine E. Mallett, Sergio F. Guarin-Perez, Alexander W. Hooke, Allison Tanner, Joshua Bland, James S. Fitzsimmons, Michael J. Taunton,Rafael J. Sierra

The Journal of Arthroplasty(2024)

引用 0|浏览0
暂无评分
摘要
Background Intra-prosthetic dissociation (IPD) is a complication unique to dual mobility (DM) implants where the outer polyethylene head dissociates from the inner femoral head. Increasing reports of IPD at the time of closed reduction of large head DM dislocations prompted this biomechanical study evaluating the assembly and dissociation forces of DM heads. Methods We tested 17 polyethylene DM heads from five vendors. Of the heads, twelve were highly cross-linked polyethylene (4 vendors), and five were infused with vitamin E (2 vendors). Heads were between 46 and 47 mm in diameter, accepting a 28 mm inner ceramic head. Implants were assembled and disassembled using a servo-hydraulic machine that recorded the forces and torques applied during testing. Dissociation was tested via both axial pull-out and lever-out techniques, where lever-out simulated stem-on-acetabular component impingement. Results The initial maximum assembly force was significantly different between all vendors (P < 0.01) and decreased for all implants with subsequent assembly. Vendor 4-E heads required the highest assembly force (1,831.9 ± 81.95 N), followed by Vendor 3, Vendor 5, Vendor 1-E, Vendor 2, and Vendor 1. Vendor 4-E implants showed the greatest dissociation resistance in both pull-out (2,059.89 N, n = 1) and lever-out (38.95 ± 2.79 Nm) tests. Vendor 1-E implants with vitamin E required higher assembly force, dissociation force, and energy than Vendor 1 heads without vitamin E. Conclusion There were notable differences in dual mobility assembly and dissociation forces between implants. Diminishing force was required for assembly with each additional trial across vendors. Vendor 4-E DM heads required the highest assembly and dissociation forces. Vitamin E appeared to increase the assembly and dissociation forces. Based on these results, DM polyethylene heads should not be reimplanted after dissociation, and there may be a role for establishing a minimum dissociation energy standard to minimize IPD risk.
更多
查看译文
关键词
Dual mobility,intra-prosthetic dissociation,dislocation,total hip arthroplasty,hip instability
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要