Accuracy of intermittently scanned continuous glucose monitoring during caesarean delivery in pregnant women with insulin-treated diabetes

Diabetes Research and Clinical Practice(2024)

引用 0|浏览3
暂无评分
摘要
Aim Continuous Glucose Monitoring (CGM) systems are not currently recommended to guide intrapartum glucose and insulin infusion, due to insufficient data. In this study, intrapartum accuracy of intermittently scanned CGM (isCGM), compared to simultaneously measured capillary glucose (CG), was evaluated. Methods Paired isCGM (Freestyle Libre 2) - CG data during caesarean delivery in pregnant women with insulin-treated diabetes were prospectively collected. The isCGM accuracy was assessed by MARD and Clarke Error Grid analysis. Moreover, the impact on intrapartum management was evaluated. Results Sixty-eight paired isCGM-CG data of 19 women were evaluated. The overallMARD was 9.28 %. All values were in A and B zones of Clarke Error Grid. Forty-six (68 %) isCGM-CG pairs were in the same glycemic range, meaning the same intrapartum management. All discordant data were identified by checking CG in case of isCGM above 110 mg/dL or less than 70 mg/dL [chi-square 21.76, p < 0.001]. At ROC curve, isCGM above 110 mg/dL was associated with 100 % sensitivity to discordant result at CG (AUC 0.859, p < 0.001). Conclusion The accuracy of isCGM during caesarean delivery was good, particularly for glucose values between 70 and 110 mg/dL, when CG confirmation could be safely avoided.
更多
查看译文
关键词
Intrapartum management,isCGM accuracy,Caesarean delivery,Diabetes in pregnancy
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要