A 7.25μK Ultra-High Resolution MEMS Resonant Thermometer

IEEE Sensors Journal(2024)

引用 0|浏览0
暂无评分
摘要
This paper proposes an ultra-high-resolution MEMS resonant thermometer that exploits differences in Young’s modulus and coefficient of thermal expansion (CTE) between structural layers to increase thermal stress due to temperature variation. The increased thermal stress eventually acts on the resonator to produce a large frequency shift, and the temperature coefficient of frequency (TCF) can be increased to 17.4 times the original, which is consistent with theoretical analysis and finite element method simulations. The MEMS resonator chip is manufactured by the standard Silicon-on-insulator process and stacked on a ceramic chip carrier through multiple layers of materials. A self-sustaining oscillator, mainly composed of a packaged MEMS resonator chip and a low-noise application-specific integrated circuit (ASIC), is built to track the resonant frequency shift of the resonator. This MEMS resonant thermometer prototype demonstrates a high temperature coefficient of frequency (TCF) of -866.84ppm/K from -50°C to 40°C and exhibits good linearity. An ultra-high resolution of 7.25μK is achieved in the closed-loop experimental test. This is the best result achieved for a MEMS thermometer employing the resonant sensing paradigm to date.
更多
查看译文
关键词
MEMS resonant thermometer,MEMS resonator,TCF,CTE,resolution