Prototype and evaluation of automatic fingertip-blood-sampling system that uses fingertip blood-vessel images to determine puncture position

Kei Takenaka, Yuji Oyamada, Masahiro Kato,Akio Nagasaka, Junpei Hokari,Takashi Irie, Nobuhiro Tsukada, Atsushi Yanagida

MEDICAL ENGINEERING & PHYSICS(2024)

引用 0|浏览0
暂无评分
摘要
We are developing an automatic fingertip-blood-sampling system to reduce the burden on trained medical personnel. For this system to withdraw a consistent volume of sampled blood for blood tests, we developed a mechanism for our system to select and puncture the vicinity of a large blood vessel from the blood-vessel image of an individual's fingertip. We call this mechanism the fingertip-vessel-puncture mechanism. From the results of an experiment in which the fingertips of 20 individuals (men and women in their 20 s to 60 s) were manually punctured at near and far locations from the blood vessel selected with our mechanism, the following conclusions were obtained. The fingertip-vessel-puncture mechanism tends to increase the volume of sampled blood, thus is effective in sampling more than 650 mu L of blood for automatic blood analyzers. It was also found that it is more effective in increasing the volume of sampled blood in the men and those who were younger.
更多
查看译文
关键词
Fingertip -blood -sampling,Automatic blood -sampling devices,Lancet,Blood -sampling tube,Fingertip -blood -vessel image