Effect of RVAD Cannulation Length on Right Ventricular Thrombosis Risk: An In Silico Investigation

Annals of Biomedical Engineering(2024)

引用 0|浏览0
暂无评分
摘要
Left ventricular assist devices (LVADs) have been used off-label as long-term support of the right heart due to the lack of a clinically approved durable right VAD (RVAD). Whilst various techniques to reduce RVAD inflow cannula protrusion have been described, the implication of the protrusion length on right heart blood flow and subsequent risk of thrombosis remains poorly understood. This study investigates the influence of RVAD diaphragmatic cannulation length on right ventricular thrombosis risk using a patient-specific right ventricle in silico model validated with particle image velocimetry. Four cannulation lengths (5, 10, 15 and 25 mm) were evaluated in a one-way fluid–structure interaction simulation with boundary conditions generated from a lumped parameter model, simulating a biventricular supported condition. Simulation results demonstrated that the 25-mm cannulation length exhibited a lower thrombosis risk compared to 5-, 10- and 15-mm cannulation lengths due to improved flow energy distribution (25.2
更多
查看译文
关键词
Right heart failure,Computational fluid dynamics (CFD),Flow modelling,Right ventricle diaphragmatic implantation,BiVAD,One-way fluid–structure interaction (FSI)