Evaluation of novel radiation protection devices during radiologically guided interventions

Maria E.V. Larsson,Pernilla I. Jonasson, Petra S. Apell,Peter P. Kearney,Charlotta J. Lundh

CVIR Endovascular(2024)

引用 0|浏览0
暂无评分
摘要
Background In radiologically guided interventions, medical practitioners are subjected to radiation exposure, which may lead to radiation-induced diseases. In this study, novel radiation shields for the head and neck were evaluated for their potential to reduce radiation exposure. Method An anthropomorphic phantom was exposed on its left side to scattered radiation from beneath to simulate the exposure of an operator in a x-ray operating room. Thermoluminescent dosimeters (TLDs) were positioned at different depths in five slices in the phantom, measuring personal dose equivalent. Two different set up situations were evaluated: a head protector designed to reduce radiation in the upper section of the head; and a novel thyroid protector prototype extended in the front and on both sides, designed to reduce radiation in the lower and middle sections of the head. A standard thyroid collar prototype and a ceiling mounted lead glass shield were used as comparisons. Furthermore, the head protector was evaluated in a clinical study in which TLDs were positioned to measure scattered radiation exposure to the heads of operators during endovascular interventions. Results The extended thyroid protector reduced the scattered radiation in the throat, chin, and ear slices. Some shielding effect was seen in the brain and skull slices. The head protector showed a shielding effect in the skull slice up to two cm depth where it covered the phantom head. As expected, the ceiling mounted lead glass shield reduced the scattered radiation in all measuring points. Conclusions A ceiling mounted lead glass shield is an effective radiation protection for the head, but in clinical practice, optimal positioning of a ceiling mounted lead shield may not always be possible, particularly during complex cases when radiation protection may be most relevant. Added protection using these novel guards may compliment the shielding effect of the ceiling mounted lead shield. The head protector stand-alone did not provide sufficient protection of the head. The extended thyroid protector stand-alone provided sufficient protection in the lower and middle sections of the head and neck. Graphical Abstract
更多
查看译文
关键词
Radiation exposure,Ionizing radiation,Scatter radiation,Radiation protection,Personal protection equipment,PPE
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要