High-throughput nano-flow proteomics using a dual-column electrospray source

An Staes,Katie Boucher, Sara Dufour, Teresa M Maia, Evy Timmerman, Delphi Van Haver, Jarne Pauwels, Hans Demol, Jonathan Vandenbussche,Francis Impens,Simon Devos

biorxiv(2024)

引用 0|浏览4
暂无评分
摘要
With current trends in proteomics, especially regarding clinical and low input (to single cell) samples, it is increasingly important to both maximize the throughput of the analysis and maintain as much sensitivity as possible. The new generation of mass spectrometers (MS) are taking a huge leap in sensitivity, allowing to analyze samples with shorter liquid chromatography (LC) methods while digging as deep in the proteome. However, the throughput can be doubled by implementing a dual column nano-LC-MS configuration. For this purpose, we used a dual-column setup with a two-outlet electrospray source, and compared it to a classic dual-column setup with a single-outlet source. ### Competing Interest Statement The authors have declared no competing interest.
更多
查看译文
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要