A calibration model transfer method based on quadrilinear model for excitation-emission matrix fluorescence measurements on different fluorescence spectrophotometers

MICROCHEMICAL JOURNAL(2024)

引用 0|浏览2
暂无评分
摘要
In this work, a calibration model transfer method based on quadrilinear model for excitation-emission matrix fluorescence (EEM) data measurements on different fluorescence spectrophotometers was proposed. The proposed method can realize the calibration model transfer of EEM data on different fluorescence spectrophotometers through only one modeling, provided that the shape of the fluorescence spectrum of each target analyte in these data does not change significantly. The calibration sample data acquired on the 2nd or 3rd instrument was transposed to that obtained from the 1st instrument using the calibration model transfer. Subsequently, the transferred data was used to build a calibration model and it was employed to predict the results of spiked prediction samples obtained from the 1st instrument. The reliability of the method was demonstrated through the simulated EEM data and the real EEM data that came from three different fluorescence spectrophotometers (HITACHI F-7000, Edinburgh FS5 and HITACHI F-4600). In addition, when unknown interference was accidentally introduced during the measurement process of calibration samples, this method support the elimination of interference before implementing AQLD and subsequent calibration model transfer was not affected. The obtained EEM data after the calibration model transfer could be combined with the second-order calibration algorithm for the quantitative analysis of target analytes in complex systems.
更多
查看译文
关键词
Calibration model transfer,Excitation -emission matrix fluorescence,Quadrilinear component model,Alternating quadrilinear decomposition,algorithm,Second -order advantage
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要