Development and establishment of a parallel plate ionization chamber as a transfer standard for measurement of dose to tissue in beta radiation fields

S. Rakshit,M.S. Kulkarni

Applied Radiation and Isotopes(2024)

引用 0|浏览0
暂无评分
摘要
A parallel-plate ionization chamber (PPC) with a nominal volume of 8.16 cm³ was developed based on theoretically simulated design parameters. Its purpose is to serve as a transfer standard for dosimetry in a beta radiation field. The entrance window of the PPC consists of an aluminized Mylar sheet with a thickness of 1.4 mg/cm2. The collecting and guard electrodes are created by applying a graphite coating on a Poly Methyl Methacrylate (PMMA) substrate with a thickness of 5 mm. The nominal sheet resistance of the graphite-coated PMMA substrate was measured using a four-probe technique and found to be approximately 800 ohms per square (Ω/□). Dosimetric characterization of the PPC was performed in the ISO 6980 reference beta radiation field, utilizing 90Sr-90Y and 85Kr beta radiation sources. The assessment included studies on short-term stability, linearity, current-to-voltage characteristics, stabilization time, and leakage current. The PPC was calibrated and established as a transfer standard using the ‘Extrapolation Ionization Chamber,' recognized as an absolute standard for dose to tissue in 90Sr-90Y and 85Kr beta sources within the laboratory. The calibration coefficient of the PPC indicates an energy dependence of 0.6 % for 90Sr-90Y and 85Kr beta sources.