High-speed Beveled Tip Versus Standard Tip Vitrectomy Probe: A Prospective Randomized Clinical Trial.

Shriji Patel,Archana Nair, Kenneth Taubenslag,Kurt Scavelli,Paul Mallory, Tomas Moreno, Rishabh Date, Heather Tamez

Journal of Ophthalmic & Vision Research(2023)

引用 0|浏览0
暂无评分
摘要
Purpose:To compare the efficiency of the advanced ultravit beveled vitrector probe (10,000 cuts per minute) to the current standard ultravit highspeed (7500 cuts per minute) vitrector probe. Methods:A prospective, randomized controlled trial was conducted on patients undergoing routine vitrectomy surgery for epiretinal membrane, full-thickness macular hole, and vitreous opacities. Patients were randomly assigned to undergo PPV with the ultravit highspeed probe (Probe 1) or the advanced ultravit beveled probe (Probe 2). The main outcome measure was time to completion of core vitrectomy and vitreous base shave. Results:Forty patients were enrolled in this study, 20 in each cohort. The average time to completion of core vitrectomy was 10.4 +/- 1.8 min in the Probe 1 cohort compared to 9.7 +/- 2 min in the Probe 2 cohort (P = 0.21). The average time to completion of vitreous base shave was 9.6 +/- 2.7 min in the Probe 1 cohort compared to 9.4 +/- 1.8 min in the Probe 2 cohort (P = 0.39). Conclusion:In the current study, the advanced ultravit beveled probe was noninferior to the ultravit highspeed vitrectomy probe when looking at the time to completion of core vitrectomy and vitreous base shave. The increased cut rate did not affect the efficiency of vitreous removal.
更多
查看译文
关键词
retina,surgery,small-gauge vitrectomy,advanced ultravit,beveled probe
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要