Left bundle branch pacing lead for sensing ventricular arrhythmias in implantable cardioverter-defibrillator: A pilot study (LBBP-ICD study).

Heart rhythm(2023)

引用 0|浏览1
暂无评分
摘要
BACKGROUND:Left bundle branch pacing (LBBP) has been suggested as an alternative modality for biventricular pacing in cardiac resynchronization therapy (CRT)-eligible patients. As it provides stable R-wave sensing, LBBP has been recently used to provide sensing of ventricular arrhythmia in patients receiving implantable cardioverter-defibrillator (ICD) with CRT. OBJECTIVE:The aim of this study was to analyze the long-term safety and efficacy of the LBBP lead for appropriate detection of ventricular arrhythmia and delivery of antitachycardia pacing (ATP) in patients requiring defibrillator therapy with CRT. METHODS:CRT-eligible patients who underwent successful LBBP-optimized ICD and LBBP-optimized CRT with defibrillator were enrolled. The LBBP lead was connected to the right ventricular-P/S port after capping the IS-1 connector plug of the DF-1-ICD lead. LBBP-optimized ICD or LBBP-optimized CRT with defibrillator was decided on the basis of correction of conduction system disease. Documented arrhythmic episodes and therapy delivered were analyzed. RESULTS:Thirty patients were enrolled. The mean age was 59.7 ± 10.5 years. LBBP resulted in an increase in left ventricular ejection fraction from 29.9% ± 4.6% to 43.9% ± 11.2% (P < .0001). During a mean follow-up of 22.9 ± 12.5 months, 254 ventricular arrhythmic events were documented. Appropriate events (n = 225 [89%]) included nonsustained ventricular tachycardia (VT) (n = 212 episodes [94%]), VT (n = 8 [3.5%]), and ventricular fibrillation (n = 5 [2.5%]). ATP efficacy in terminating VT was 75%. Eleven percent of episodes (n = 29) were inappropriately detected because of T-wave oversensing. Inappropriate therapy (ATP) was delivered for 14 episodes (5.5%). Three patients (10%) had worsening of tricuspid regurgitation. CONCLUSION:Sensing from the LBBP lead for arrhythmia detection is safe as ∼90% of the episodes were detected appropriately. Future studies with a dedicated LBBP-defibrillator lead along with algorithms to avoid oversensing can help in combining defibrillation with conduction system pacing.
更多
查看译文
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要