Eight-channel SiNx microring-resonator based photonic biosensor for label-free fluid analysis in the optical C-band.

Jakob Reck,Moritz Kleinert,Klara Mihov,Martin Kresse, Cafercan Yilmaz,Christina Hoffmann, Peter Hoffmann, Vera Froese, Ulrich Kertzscher, Kristina Mykhailiuk, Julia Michaelis, Wilfried Weigel, Soren Scholand, Hans-Jurgen Heupke,Madeleine Weigel,David De Felipe,Tianwen Qian,Hauke Conradi,Crispin Zawadzki,Norbert Keil,Martin Schell

2023 45th Annual International Conference of the IEEE Engineering in Medicine & Biology Society (EMBC)(2023)

引用 0|浏览12
暂无评分
摘要
A lab-on-a-chip multichannel sensing platform for biomedical analysis based on optical silicon nitride (SiNx) microring-resonators (MRR) was established. The resonators were surface functionalized and finally combined with a microfluidic chamber for validation using an avidin-biotin ligand-binding assay. The results with a limit of detection (LOD) of 2.3∙10-5 and a mean intra-assay coefficient of variation (CV) of ±10.0 %, also under consideration of FDA guidelines, show promising future applicability for a wide variety of targets in the field of outpatient medical diagnostics and life science.Clinical Relevance- Biomarkers play a crucial role in physiological processes of the human body. To enable instantaneous and decentralized analysis of these markers, systems are needed that can be used in a laboratory-independent environment with minimal amounts of biofluid. An example is the utilization of such systems for neonates or infants.