Characterization of mechanical and radiation isocenter on an MR‐guided radiotherapy (MRgRT) Linac

Journal of Applied Clinical Medical Physics(2023)

引用 0|浏览0
暂无评分
摘要
Abstract Background and Purpose In the emerging paradigm of stereotactic radiosurgery being proposed for MR‐guided radiotherapy (MRgRT), assessment of mechanical geometric accuracy is critical for the implementation of stereotactic delivery. We benchmarked the mechanical accuracy of an MR Linac system that lacks an onboard detector/array. Our mechanical tests utilize a half beam block (HBB) geometry that takes advantage of the sensitivity of a partially occluded detector. Materials and Methods Mechanical tests benchmarked the couch, MLC, and gantry geometric accuracy for an MR‐Linac system. An HBB technique was used to irradiate an ionization chamber profiler (ICP) array with partial occlusion of individual detectors for characterization of MLC skew, beam divergence displacement, and RT isocenter localization. The sensitivity of the partially occluded detector's ICP‐X (detector width) and ICP‐Y (detector length) was characterized by displacing the detector relative to radiation isocenter by 0.2 mm increments, introduced through couch motion. The accuracy of the HBB ICP technique was verified with a starshot using radiochromic film, and the reproducibility was verified on a conventional C‐arm Linac and compared to Winston‐Lutz. Results The sensitivity of the HBB technique as quantified through the dose difference normalized to open field as a function of displacement from RT isocenter was 6.4%/mm and 13.0%/mm for the ICP‐X and ICP‐Y orientation, respectively, due to the oblong detector orientation. Couch positional accuracy and sag was within ±0.1 mm. Maximum MLC positional displacement was 0.7 mm with mean MLC skew at 0.07°. The maximum beam divergence displacement was 0.03 mm. The gantry angle was within 0.1°. Independent verification of the RT isocenter localization procedure produced repeatable results. Conclusion This work serves for characterizing the mechanical and geometric radiation accuracy for the foundation of an MR‐guided stereotactic radiosurgery program, as demonstrated with high sensitivity and independent validation.
更多
查看译文
关键词
radiation isocenter,radiotherapy,mrgrt
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要