Cycle Threshold (Ct) Values of SARS-CoV-2 Detected with the GeneXpert® System and a Mutation Associated with Different Target Gene Failure

Current Issues in Molecular Biology(2023)

引用 0|浏览3
暂无评分
摘要
SARS-CoV-2 nucleic acid detection tests enable rapid virus detection; however, it is challenging to identify genotypes to comprehend the local epidemiology and infection routes in real-time qRT-PCR. At the end of June 2022, our hospital experienced an in-hospital cluster of COVID-19. When examined using the GeneXpert® System, the cycle threshold (Ct) value of the N2 region of the nucleocapsid gene of SARS-CoV-2 was approximately 10 cycles higher than that of the envelope gene. Sanger sequencing revealed a G29179T mutation in the primer and probe binding sites. A review of past test results revealed differences in Ct values in 21 of 345 SARS-CoV-2-positive patients, of which 17 cases were cluster-related and 4 were not. Including these 21 cases, 36 cases in total were selected for whole-genome sequencing (WGS). The viral genomes in the cluster-related cases were identified as BA.2.10, and those in the non-cluster cases were closely related and classified as being downstream of BA.2.10 and other lineages. Although WGS can provide comprehensive information, its use is limited in various laboratory settings. A measurement platform reporting and comparing Ct values of different target genes can improve test accuracy, enhance our understanding of infection spread, and be applied to the quality control of reagents.