Reduced Iodinated Contrast Media Administration in Coronary CT Angiography on a Clinical Photon-Counting Detector CT System

Investigative Radiology(2022)

引用 20|浏览1
暂无评分
摘要
Purpose The aim of this study was to evaluate strategies to reduce contrast media volumes for coronary computed tomography (CT) angiography on a clinical first-generation dual-source photon-counting detector (PCD)-CT system using a dynamic circulation phantom. Materials and Methods Coronary CT angiograph is an established method for the assessment of coronary artery disease that relies on the administration of iodinated contrast media. Reduction of contrast media volumes while maintaining diagnostic image quality is desirable. In this study, a dynamic phantom containing a 3-dimensional-printed model of the thoracic aorta and coronary arteries was evaluated using a clinical contrast injection protocol with stepwise reduced contrast agent concentrations (100%, 75%, 50%, 40%, 30%, and 20% contrast media content of the same 50 mL bolus, resulting in iodine delivery rates of 1.5, 1.1, 0.7, 0.6, 0.4 and 0.3 gl/s) on a first-generation, dual-source PCD-CT. Polychromatic images (T3D) and virtual monoenergetic images were reconstructed in the range of 40 to 70 keV in 5-keV steps. Attenuation and noise were measured in the coronary arteries and background material and the contrast-to-noise ratio (CNR) were calculated. Attenuation of 350 HU and a CNR of the reference protocol at 70 keV were regarded as sufficient for simulation of diagnostic purposes. Vessel sharpness and noise power spectra were analyzed for the aforementioned reconstructions. Results The standard clinical contrast protocol (bolus with 100% contrast) yielded diagnostic coronary artery attenuation for all tested reconstructions (>398 HU). A 50% reduction in contrast media concentration demonstrated sufficient attenuation of the coronary arteries at 40 to 55 keV (>366 HU). Virtual monoenergetic image reconstructions of 40 to 45 and 40 keV allowed satisfactory attenuation of the coronary arteries for contrast concentrations of 40% and 30% of the original protocol. A reduction of contrast agent concentration to 20% of the initial concentration provided insufficient attenuation in the target vessels for all reconstructions. The highest CNR was found for virtual monoenergetic reconstructions at 40 keV for all contrast media injection protocols, yielding a sufficient CNR at a 50% reduction of contrast agent concentration. Conclusions Using virtual monoenergetic image reconstructions at 40 keV on a dual-source PCD-CT system, contrast media concentration could be reduced by 50% to obtain diagnostic attenuation and objective image quality for coronary CT angiography in a dynamic vessel phantom. These initial feasibility study results have to be validated in clinical studies.
更多
查看译文
关键词
coronary ct angiography,iodinated contrast media administration,photon-counting
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要