Neurointerventions on two generations of angiography systems: Recent systems reduce radiation exposure by half

Physica Medica(2024)

引用 0|浏览9
暂无评分
摘要
•Air kerma at the reference point (Ka,r) on a new generation of angiography system was 48% of an older reference system.•Air kerma-area product (KAP) on the new system was 50% of the older system.•Substantial decrease in radiation exposure to the patients may also translate to the staff.
更多
查看译文
关键词
FGI,FWHM,HVL,Kar,KAP
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要