Transcatheter Aortic Valve Migration During Vascular Access Closure

CUREUS JOURNAL OF MEDICAL SCIENCE(2023)

引用 0|浏览2
暂无评分
摘要
We present a novel complication of transcatheter aortic valve replacement (TAVR) involving prosthetic migration due to entanglement by a standard guidewire during non-large bore vascular access closure, followed by successful bail-out using a second transcatheter prosthesis. To knowledge, this mechanism of prosthesis migration has not been previously described.
更多
查看译文
关键词
tavr,perclose proglide,iatrogenic complication,prosthetic valve migration,transcatheter aortic valve