Fractal capacitive electrodes have been successfully integrated into a digital microfluidic open-platform. These electrodes perform actuation and withstand voltages up to 300V without insulation-layer breakdown. They were used to quantify the concentration levels of C-reactive protein (CRP) to determine the risk of cardiovascular disease. The capacitance increased sevenfold and stabilized in less than 5 minutes. The sensor shows a decreasing trend of capacitance readouts with the increase of concentrations. The same immunoassay was tested with untreated electrodes and showed no significant response, which suggests that immobilization was necessary. This configuration allows the electrodes to be used as biosensors.
更多
查看译文
关键词
Fractal Capacitor,Biosensor,Digital Microfluidic,EWOD,Electrowetting On Dielectric,C-Reactive Protein,CRP