Novel Deep Learning Denoising Enhances Image Quality and Lowers Radiation Exposure in Interventional Bronchial Artery Embolization Cone Beam CT.

Andreas S Brendlin, Reza Dehdab, Benedikt Stenzl, Jonas Mueck,Patrick Ghibes, Gerd Groezinger,Jonghyo Kim,Saif Afat,Christoph Artzner

Academic radiology(2023)

引用 0|浏览0
暂无评分
摘要
OBJECTIVES:In interventional bronchial artery embolization (BAE), periprocedural cone beam CT (CBCT) improves guiding and localization. However, a trade-off exists between 6-second runs (high radiation dose and motion artifacts, but low noise) and 3-second runs (vice versa). This study aimed to determine the efficacy of an advanced deep learning denoising (DLD) technique in mitigating the trade-offs related to radiation dose and image quality during interventional BAE CBCT. MATERIALS AND METHODS:This study included BMI-matched patients undergoing 6-second and 3-second BAE CBCT scans. The dose-area product values (DAP) were obtained. All datasets were reconstructed using standard weighted filtered back projection (OR) and a novel DLD software. Objective image metrics were derived from place-consistent regions of interest, including CT numbers of the Aorta and lung, noise, and contrast-to-noise ratio. Three blinded radiologists performed subjective assessments regarding image quality, sharpness, contrast, and motion artifacts on all dataset combinations in a forced-choice setup (-1 = inferior, 0 = equal; 1 = superior). The points were averaged per item for a total score. Statistical analysis ensued using a properly corrected mixed-effects model with post hoc pairwise comparisons. RESULTS:Sixty patients were assessed in 30 matched pairs (age 64 ± 15 years; 10 female). The mean DAP for the 6 s and 3 s runs was 2199 ± 185 µGym² and 1227 ± 90 µGym², respectively. Neither low-dose imaging nor the reconstruction method introduced a significant HU shift (p ≥ 0.127). The 3 s-DLD presented the least noise and superior contrast-to-noise ratio (CNR) (p < 0.001). While subjective evaluation revealed no noticeable distinction between 6 s-DLD and 3 s-DLD in terms of quality (p ≥ 0.996), both outperformed the OR variants (p < 0.001). The 3 s datasets exhibited fewer motion artifacts than the 6 s datasets (p < 0.001). CONCLUSIONS:DLD effectively mitigates the trade-off between radiation dose, image noise, and motion artifact burden in regular reconstructed BAE CBCT by enabling diagnostic scans with low radiation exposure and inherently low motion artifact burden at short examination times.
更多
查看译文
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要