Initial clinical experience of atrial fibrillation ablation guided by a cryoballoon-compatible, magnetic-based circular catheter

JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY(2024)

引用 0|浏览8
暂无评分
摘要
IntroductionThe circular catheter compatible with current cryoballoon system for atrial fibrillation (AF) ablation is exclusively sensed by impedance-based electro-anatomical mapping (EAM) system, limiting the accuracy of maps. We aim to investigate the feasibility and safety of a magnetic-based circular mapping catheter for AF ablation with cryoballoon.MethodsNineteen consecutive patients who underwent pulmonary vein isolation (PVI) with cryoballoon for paroxysmal or persistent AF were included. EAMs of left atrium (LA) created by the LASSOSTAR (TM) NAV catheter (Lassostar map) before and after PVI were compared to that generated by a high-density mapping catheter (Pentaray map) from different aspects including structural similarity, PV angle, LA posterior wall (LAPW) and low voltage areas (LVAs), and the amplitude of far field electrograms (FFEs) recorded by catheters.ResultsAll patients had successful PVI without major complications. With similar mapping time and density, the LA volume calculated from the Pentaray map and Lassostar map were comparable. There were no significant differences in PV angle of all PVs and PW area (16.8 +/- 3.2 vs. 17.1 +/- 2.8, p = .516) between Pentaray map and Lassostar map. High structural similarity score was observed between two maps (0.783 in RAO/LAO view and 0.791 in PA view). Lassostar map detected lesser but not statistically significant extension of LVA (13.9% vs. 18.3%, p = .07). Amplitude of FFE was larger at the right superior PV on Lassostar map (0.21 +/- 0.16 vs. 0.14 +/- 0.11 mV, p = .041) compared to that on the Pentaray map.ConclusionIn our initial experience, PVI with cryoballoon and magnetic-based circular LASSOSTAR (TM) NAV catheter was safe and effective based on the accurate LA geometry it created.
更多
查看译文
关键词
ablation,atrial fibrillation,circular catheter,cryoballoon,electro-anatomical mapping,pulmonary vein isolation
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要