Increased stability due to symmetric cement volume in augmented pedicle screws? A biomechanical study.

Journal of biomedical materials research. Part B, Applied biomaterials(2023)

引用 0|浏览5
暂无评分
摘要
Pedicle screw instrumentation has become "state of the art" in surgical treatment of many spinal disorders. Loosening of pedicle screws due to poor bone mineral density is a frequent complication in osteoporotic patients. As prevalence of osteoporosis and spinal disorders are increasing with an aging demographic, optimizing the biomechanical properties of pedicle screw constructions and therefore outcome after spinal surgery in osteoporotic patients is a key factor in future surgical therapy. Therefore, this biomechanical study investigated the stability of polymethylmethacrylate (PMMA)-augmented pedicle screw-rod constructions under a deviating distribution of PMMA applied to the instrumentation in osteoporotic human cadaveric vertebrae. We showed that PMMA-augmented pedicle screw-rod constructions tend to be more stable than those with non-augmented pedicle screws. Further, there appears to be a larger risk of screw loosening in unilateral augmented pedicle screws than in non-augmented, therefore a highly asymmetrically distributed PMMA should be avoided.
更多
查看译文
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper