Radionuclide-free efficiency calibration of an HPGe detector using monochromatic photon beams calibrated with a cryogenic radiometer

APPLIED RADIATION AND ISOTOPES(2024)

引用 0|浏览0
暂无评分
摘要
A cryogenic electrical-substitution radiometer was used for the primary calibration of monochromatic photon beams at synchrotron, which are in turn employed to determine the intrinsic efficiency of an HPGe detector between 3.5 keV and 8 keV. This allowed the determination of the thickness of the dead layer, which enabled the calculation of the efficiency up to 55 keV, where the active thickness is not determinant. The geometrical efficiency was determined through the measurement of a radioactive source at variable distances.
更多
查看译文
关键词
Semiconductor detector,Efficiency calibration,Photon emission intensities,Cryogenic detector,Synchrotron radiation