Transcranial dipole localization and decoding study based on ultrasonic phased array for acoustoelectric brain imaging.

Journal of neural engineering(2023)

引用 0|浏览6
暂无评分
摘要
Objective. Neuroimaging is one of the effective tools to understand the functional activities of the brain, but traditional non-invasive neuroimaging techniques are difficult to combine both high temporal and spatial resolution to satisfy clinical needs. Acoustoelectric brain imaging (ABI) can combine the millimeter spatial resolution advantage of focused ultrasound with the millisecond temporal resolution advantage of electroencephalogram signals.Approach. In this study, we first explored the transcranial modulated acoustic field distribution based on ABI, and further localized and decoded single and double dipoles signals.Main results. The results show that the simulation-guided acoustic field modulation results are significantly better than those of self-focusing, which can realize precise modulation focusing of intracranial target focusing. The single dipole transcranial localization error is less than 0.4 mm and the decoding accuracy is greater than 0.93. The double dipoles transcranial localization error is less than 0.2 mm and the decoding accuracy is greater than 0.89.Significance. This study enables precise focusing of transcranial acoustic field modulation, high-precision localization of source signals and decoding of their waveforms, which provides a technical method for ABI in localizing evoked excitatory neuron areas and epileptic focus.
更多
查看译文
关键词
transcranial dipole localization,brain imaging
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper