A new system for in-room image guidance in particle therapy at CNAO.

Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB)(2023)

引用 0|浏览1
暂无评分
摘要
This paper describes the design, installation, and commissioning of an in-room imaging device developed at the Centro Nazionale di Adroterapia Oncologica (CNAO, Pavia, Italy). The system is an upgraded version of the one previously installed in 2014, and its design accounted for the experience gained in a decade of clinical practice of patient setup verification and correction through robotic-supported, off-isocenter in-room image guidance. The system's basic feature consists of image-based setup correction through 2D/3D and 3D/3D registration through a dedicated HW/SW platform. The major update with respect to the device already under clinical usage resides in the implementation of a functionality for extending the field of view of the reconstructed Cone Beam CT (CBCT) volume, along with improved overall safety and functional optimization. We report here details on the procedures implemented for system calibration under all imaging modalities and the results of the technical and preclinical commissioning of the device performed on two different phantoms. In the technical commissioning, specific attention was given to the assessment of the accuracy with which the six-degrees-of-freedom correction vector computed at the off-isocenter imaging position was propagated to the planned isocentric irradiation geometry. During the preclinical commissioning, the entire clinical-like procedure for detecting and correcting imposed, known setup deviation was tested on an anthropomorphic radioequivalent phantom. Results showed system performance within the sub-millimeter and sub-degree range according to project specifications under each imaging modality, making it ready for clinical application.
更多
查看译文
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要