Novel production method for traceable surface sources by aluminium functionalisation.

D Tüzün, L Chambon, Y Kergadallan,V Lourenço

Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine(2023)

引用 0|浏览0
暂无评分
摘要
Accurate detection of low-level radioactivity is critical in decommissioning. However, commercial sources used for calibration lack representativeness due to their flat surface. The objective of this work is to produce flexible and large area surface sources for alpha and beta emitters by functionalising aluminium foil. Functionalisation strategies were developed to provide these sources. The manganese oxide coated sample shows the highest fixation yield for Am: (70.0 ± 2.6)% and conforms to ISO 8769 for uniformity (92%).
更多
查看译文
关键词
Radioactive source,Large area surface sources,Surface functionalisation,Eu-152,Am-241