Analytical and clinical evaluations of SNIBE Maglumi chemiluminescent immunoassay for the detection of SARS-CoV-2 antigen in salivary samples

CLINICAL CHEMISTRY AND LABORATORY MEDICINE(2024)

引用 0|浏览10
暂无评分
摘要
Objectives: In this study, we describe the analytical and clinical performances of the SNIBE Maglumi SARS-CoV-2 antigen fully-automated chemiluminescent immunoassay (MAG-CLIA) on salivary samples.Methods: Limit of detection (LOD), linearity and precision were tested for values close to or below the declared LOD. Clinical performance of MAG-CLIA was evaluated on leftover salivary samples from the healthcare workers (HCW) surveillance program, at the University-Hospital of Padova. Salivary samples were analyzed by Lumipulse G SARS-CoV-2 Ag, and in case where the values exceeded 0.41 ng/L, further testing was conducted using TaqPathTM COVID-19 RT-PCR (Applied Biosystems, Thermo Fisher Scientific).Results: The estimated MAG-CLIA LOD was 3 ng/L, with repeatability of 7.5 %. Good linearity was demonstrated by diluting two samples at 52.7 ng/L and 211.4 ng/L. Of the 228 HCW samples, 59/228 (25.9 %) were positive, 169/228 (74.1 %) were negative. MAG-CLIA SARS-CoV-2 sAg median level (and interquartile range [IQR]) was 5.03 ng/L (<0.001-35.8 ng/L) for positive and <0.001 ng/L (<0.001 ng/L) for negative samples. MAG-CLIA AUC was 0.795 (95 % CI: 0.720-0.871). Using the best cut-off, 3.5 ng/L, sensitivity and specificity were 57.1 % (95 % CI: 42.2-71.2 %) and 97.0 % (95 % CI: 93.2-99.0 %), respectively. The agreement with the molecular assay was 88.1 % (Cohen's kappa 0.606 [SE=0.066, p<0.001]).Conclusions: The analytical performances of MAG-CLIA are satisfactory, also when values below LOD were tested. In saliva samples, although specificity was elevated, clinical performance was not comparable with that on nasopharyngeal swabs (NPS).
更多
查看译文
关键词
SARS-CoV-2 antigen,saliva,COVID-19,chemiluminescent assay,Maglumi
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要