Modernization of safety environment for a dedicated beamline for proton ocular therapy.

Medical physics(2023)

引用 0|浏览4
暂无评分
摘要
Throughout the task of modernizing the safety system of our dedicated ocular beamline, FMEA proved to be an effective instrument in soliciting inputs from the staff about safety and workflow concerns, helping to identify steps associated with elevated failure risks. Risks were reduced with the clinical introduction of a new IGRT application, which integrates quality management tools widely recognized for their role in risk mitigation: automation of the data transfer and workflow steps, and with the introduction of checklists and redundancy cross-checks.
更多
查看译文
关键词
failure mode and effect analysis, image-guided radiation therapy, ocular cancers, proton therapy