Evaluation of the Catheter Clamp over Hydrophilic Guide Wire Central Venous Catheter Exchange Technique for Air Embolism Prophylaxis in an In Vitro Model

JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY(2024)

引用 0|浏览4
暂无评分
摘要
Purpose: To develop a reproducible in vitro model simulating central venous catheter (CVC) exchange with high potential for air embolization and test the hypothesis that a closed catheter clamp over hydrophilic guide wire exchange technique will significantly reduce the volume of air introduced during CVC exchange. Materials and Methods: The model consisted of a 16-F valved sheath, 240-mL container, and pressure transducer submerged in water in a 1,200-mL suction canister system. Continuous wall suction was applied to the canister to maintain negative pressure at -7 mm Hg or -11 mm Hg. Each trial consisted of 0.035-inch hydrophilic guide wire introduction, over-the-wire catheter exchange, and wire removal following clinical protocol. A total of 256 trials were performed, 128 trials at each pressure with the catheter clamp open (n = 64) or closed (n = 64) around the hydrophilic guide wire. Results: There was a statistically significant lower volume of air introduced with closed clamp over-the-wire exchanges than with open clamp exchanges at both pressures (2-tailed t-test, P < .001). At -7 mm Hg, a mean of 48.0 mL (SD +/- 9.3) of air was introduced with open clamp and 20.6 mL (SD +/- 4.7) of air was introduced with closed clamp. At -11 mm Hg, 97.8 mL (SD +/- 11.9) of air was introduced with open clamp and 37.8 mL (SD +/- 6.3) of air was introduced with closed clamp. Conclusions: This study demonstrated the use of a reproducible in vitro model mimicking conditions causing air embolism during CVC exchange. Results showed that CVC exchange using closed catheter clamp over hydrophilic guide wire exchange technique significantly reduced the volume of air introduced per exchange.
更多
查看译文
关键词
AE,CVC,HD
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要