Evaluation of lesion characteristics and baseline impedance on high-power short-duration radiofrequency catheter ablation using computer simulation

Yao Sun,Xin Zhu,Keijiro Nakamura, Shuyu Wang

Heart and vessels(2023)

引用 0|浏览4
暂无评分
摘要
Myocardium baseline impedance (BI) is an important factor in ablation effectiveness. This study examined the performance of low-power and long-duration (LPLD), high-power and short-duration (HPSD) ablation at different BIs by computer simulation. A 3D model of the ablation region was constructed for simulation, and in vitro experiments were performed to validate the simulation. Three ablation power and duration configurations of 30 W/30 s, 50 W/10 s, and 90 W/5 s were used for simulation with BI values of 90, 100, 110, 120, 130, and 140 Ω. Roll-off time and ablation volume were measured to evaluate ablation results. The simulation is consistent with the in vitro experiments. When BI is changed from 90 Ω to 140 Ω , the lesion volume over 50 °C with BI of 140 Ω was reduced by 6.3
更多
查看译文
关键词
baseline impedance,high-power,short-duration
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要