The population percentile allowance method for determining systematic spatial error tolerances for temporary intensity modulated brachytherapy.

Medical physics(2023)

引用 0|浏览10
暂无评分
摘要
The PPA method was formalized to determine the temporary IMBT spatial uncertainty tolerances that would be expected to result in an allowed percentage of a population of patients receiving relative dose-volume errors above a defined percentage. Multi-shield, helical Yb-based RSBT for cervical cancer was evaluated and tolerances determined, which, if applied on each treatment fraction, would represent an extreme situation. The PPA method is applicable to a variety of temporary IMBT approaches and can be used to rigorously determine the design parameters for the delivery systems such as mechanical driver motor accuracy, shield angle backlash, applicator rotation, and applicator fixation stability.
更多
查看译文
关键词
brachytherapy delivery uncertainty, DMBT, dynamic modulated brachytherapy, error tolerance, HDR, high-dose-rate brachytherapy, IMBT, Intensity modulated brachytherapy, patient population allowance, PPA, rotating shield brachytherapy, RSBT, temporary IMBT