Establishment and study of a polarized X-ray radiation facility

chinaxiv(2023)

引用 0|浏览8
暂无评分
摘要
With the advancement in X-ray astronomical detection technology, various celestial polarization detection projects have been initiated. To meet the calibration requirements of polarimeters on the ground, a polarized X-ray radiation facility was designed for this study. The design was based on the principle that X-rays incident at 45$^\circ$ on a crystal produce polarized X-rays, and a second crystal was used to measure the polarization of the X-rays produced by the facility after rotation. The effects of different diaphragm sizes on the degree of polarization were compared, and the facility produced X-rays with polarization degrees of up to 99.55$\pm$0.96\% using LiF200 and LiF220 crystals. This result revealed that the polarization of incident X-rays is one of the factors affecting the diffraction efficiency of crystals. The replacement of different crystals can satisfy the calibration requirements of polarized X-ray detectors with more energy points in the energy range (4–10) keV. In the future, the facility should be placed in a vacuum environment to meet the calibration requirements at lower energies.
更多
查看译文
关键词
: Polarized X-rays,Polarimetry,Calibration,Bragg diffraction