Design and evaluation of the pneumatic leg prosthesis ERiK to assist elderly amputees with sit-down and stand-up movements

Heike Vallery, Frederik Lachmann, Simon van der Helm, Andrew Pennycott,Gerwin Smit

WEARABLE TECHNOLOGIES(2023)

引用 0|浏览0
暂无评分
摘要
Standing up using one leg is a challenging task for those with a transfemoral amputation, particularly for elderly users with a low activity level. Active prostheses are generally not accessible to this group and available passive prostheses do not support standing up. This article presents the design and evaluation of the "Energy Restoring Intelligent Knee" (ERiK), which stores energy during sit-down in a pneumatic cylinder and returns it during stand-up. We hypothesized that the system would reduce the time needed to perform transitions and also enable higher load sharing by the prosthetic leg. However, the results of an experimental study with seven participants with transfemoral amputation contradict these hypotheses: the participants could neither move faster nor make more use of the prosthetic leg to share their body weight during transitions. We observed that a major obstacle to the useful functionality of the leg was the absence of ankle dorsiflexion - the foot tended to slip during stand-up initiation, such that only low pre-pressures and therefore support levels could be set. The rather binary action of the pneumatics also complicated movement initiation. The lessons learned from this study may be helpful to those seeking to create better designs in the future.
更多
查看译文
关键词
Knee exoprosthesis,sit-to-stand,lightweight pneumatic actuator,transfemoral amputation
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要