Proposing an intelligent technique based on radial basis function neural network to forecast the energy spectrum of diagnostic X-ray imaging systems

Zhanjian Cai, Jiadi Zheng, Liu Shan,Wang Wei, Wenzong Zhu,Yanjie Lu, Xicai Zhang, Guoqiang Xu

APPLIED RADIATION AND ISOTOPES(2023)

引用 0|浏览1
暂无评分
摘要
In digital subtraction angiography (digital subtraction total cerebral angiography), cardiac and macrovascular cardiography, anorectal radiology, fluoroscopy, and computed tomography (CT), a prior knowledge to X-ray energy spectrum is crucial for assessing the image quality and also calculating patient X-ray dosage. The present investigation's main objective is to propose an intelligent technique for faster calculating X-ray energy spectrum of medical imaging systems with different exposure settings of tube voltage, filter material, and thickness based on limited specific spectra. In this study, Monte Carlo N Particle (MCNP) simulation code was initially used to generate some limited X-ray spectra for tube voltages of 20, 30, 40, 50, 80, 100, 130, and 150 kV for two different filters of beryllium and aluminum with thicknesses of 0. 4, 0.8, 1.2, 1.6 and 2 mm. Tube voltage, type, and thickness of filter were used as the 3 inputs of 150 Radial Basis Function Neural Network (RBFNN) to forecast point by point of the X-ray spectrum. After training, the RBFNNs could forecast most of the X-ray spectra for tube voltages in the range of 20-150 kV and two various filters of aluminum and beryllium with thicknesses in the range of 0-2 mm.
更多
查看译文
关键词
Energy spectra,Diagnostic,Radial basis function neural network
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要