Prevalence and management of electrical lead abnormalities in cardiac implantable electronic device leads.

Heart rhythm O2(2023)

引用 0|浏览10
暂无评分
摘要
Background:Electrical lead abnormalities (ELAs) can result in device malfunction, leading to significant morbidity in patients with cardiac implantable electronic devices (CIEDs). Objective:We sought to determine the prevalence and management of ELAs in patients with CIEDs. Methods:This was a retrospective cohort study of patients implanted with a CIED between 2012 and 2019 at a tertiary care center. The primary outcome was ELA defined as increased capture threshold (≥2× implantation value), decreased sensing (≤0.5 implantation value), change in impedance (>50% over 3 months), or nonphysiologic potentials. A secondary outcome of device clinic utilization was also collected. Results:There were 2996 unique patients (35% female) included with 4600 leads (57% Abbott, 43% Medtronic). ELAs were observed in 135 (3%) leads, including 124 (92%) Abbott and 10 (7%) Medtronic leads (hazard ratio 9.25, P < .001). Mean follow-up was 4.5 ± 2.2 years. ELAs were associated smaller lead French size, atrial location, and Abbott leads. Lead revision was required in 28% of cases. Patients with lead abnormalities had 38% more in-clinic visits per patient year of follow-up compared with those without (P < .001). Conclusion:ELAs were more frequent in certain models, which increased rates of revision and follow-up. Identification of factors that mitigate these abnormalities to improve lead performance are required to improve care for these devices and provide efficient healthcare.
更多
查看译文
关键词
Cardiac implantable electronic devices,Lead failure,Pacing leads,Tendril leads,Electrical noise
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要