Bending Angle Sensor Based on Double-Layer Capacitance Suitable for Human Joint.

Daisuke Goto,Yusuke Sakaue, Tatsuya Kobayashi, Kohei Kawamura,Shima Okada,Naruhiro Shiozawa

IEEE Open Journal of Engineering in Medicine and Biology(2023)

引用 0|浏览0
暂无评分
摘要
Goal: To develop bending angle sensors based on double-layer capacitance for monitoring joint angles during cycling exercises. Methods: We develop a bending angle sensor based on double-layer capacitive and conducted three stretching, bending, and cycling tests to evaluate its validity. Results: We demonstrate that the bending angle sensor based on double-layer capacitance minimizes the change in the capacitance difference in the stretching test. The hysteresis and root mean square error (RMSE) compared with the optical motion capture show hysteresis: 8.0% RMSE and 3.1° in the bending test. Moreover, a cycling experiment for human joint angle measurements confirm the changes in accuracy. The RMSEs ranged from 4.7° to 7.0°, even when a human wears leggings fixed with the developed bending-angle sensor in the cycling test. Conclusion: The developed bending angle sensor provides a practical application of the quantitative and observational evaluation tool for knee joint angles.
更多
查看译文
关键词
Cycling,hysteresis,knee joint,rehabilitation,wearable sensor