Fully Endoscopic Mitral Valve Repair with Percutaneous Cannulation of Groin Vessels.

Journal of visualized experiments : JoVE(2023)

引用 0|浏览4
暂无评分
摘要
Endoscopic mitral valve surgery (EMS) has become a standard of care at specialized heart centers, further reducing surgical trauma compared to a traditional minimally invasive, thoracotomy-based approach. Exposure of the groin vessels for the establishment of cardiopulmonary bypass (CPB) via surgical cutdown in minimally invasive surgery (MIS) may result in wound healing disorders or seroma formation. The avoidance of surgical exposure of the groin vessels by using fully percutaneous techniques for the insertion of a CPB cannula with the implementation of vascular pre-closure devices has the potential to reduce these complications and improve clinical results. Herein, we present the utilization of a novel plug based vacsular closure device with a resobable collagen plug and the absence of suture material for closure of the arterial access for CPB in MIS. While this device was initially predominantly used in transcatheter aortic valve implantation (TAVI) procedures, with its safety and feasibility shown, we herein show that it can be used in CPB cannulation, since it is capable of closing arterial access sites up to 25 French (Fr.) in size. This device may be suitable to significantly reduce groin complications in MIS and simplify the establishment of CPB. Here, we describe the fundamental steps of EMS, including percutaneous groin cannulation and decannulation using a vascular closure device.
更多