3-D Ultrafast Shear Wave Absolute Vibro-Elastography Using a Matrix Array Transducer

IEEE transactions on ultrasonics, ferroelectrics, and frequency control(2023)

引用 0|浏览8
暂无评分
摘要
Real-time ultrasound imaging plays an important role in ultrasound-guided interventions. The 3-D imaging provides more spatial information compared to conventional 2-D frames by considering the volumes of data. One of the main bottlenecks of 3-D imaging is the long data acquisition time, which reduces practicality and can introduce artifacts from unwanted patient or sonographer motion. This article introduces the first shear wave absolute vibro-elastography (S-WAVE) method with real-time volumetric acquisition using a matrix array transducer. In S-WAVE, an external vibration source generates mechanical vibrations inside the tissue. The tissue motion is then estimated and used in solving a wave equation inverse problem to provide the tissue elasticity. A matrix array transducer is used with a Verasonics ultrasound machine and a frame rate of 2000 volumes/s to acquire 100 radio frequency (RF) volumes in 0.05 s. Using plane wave (PW) and compounded diverging wave (CDW) imaging methods, we estimate axial, lateral, and elevational displacements over 3-D volumes. The curl of the displacements is used with local frequency estimation to estimate elasticity in the acquired volumes. Ultrafast acquisition extends substantially the possible S-WAVE excitation frequency range, now up to 800 Hz, enabling new tissue modeling and characterization. The method was validated on three homogeneous liver fibrosis phantoms and on four different inclusions within a heterogeneous phantom. The homogeneous phantom results show less than 8% (PW) and 5% (CDW) difference between the manufacturer values and the corresponding estimated values over a frequency range of 80–800 Hz. The estimated elasticity values for the heterogeneous phantom at 400-Hz excitation frequency show the average errors of 9% (PW) and 6% (CDW) compared to the provided average values by magnetic resonance elastography (MRE). Furthermore, both imaging methods were able to detect the inclusions within the elasticity volumes. An ex vivo study on a bovine liver sample shows less than 11% (PW) and 9% (CDW) difference between the estimated elasticity ranges by the proposed method and the elasticity ranges provided by MRE and acoustic radiation force impulse (ARFI).
更多
查看译文
关键词
Transducers, Imaging, Ultrasonic imaging, Three-dimensional displays, Elasticity, Radio frequency, Liver, 3-D elastography, elasticity estimation, real-time elastography, ultrafast elastography, ultrasound
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要