Urine Dipstick Analysis on Automated Platforms: Is a Reliable Screening Tool for Proteinuria? An Experience from Umberto I Hospital in Rome.

Biomedicines(2023)

引用 0|浏览6
暂无评分
摘要
Urinalysis is commonly used as a screening tool for kidney disease. In many cases, the dipstick urine assay includes the assessment of albumin/protein and creatinine; consequently, the value of their ratio is available on the urine section report. Identification of albuminuria/proteinuria at early stages is an important issue to prevent or at least delay the onset of chronic kidney disease (CKD), kidney failure, and the progression of cardiovascular damage linked to the kidney's loss of function. Sensitive and specific diagnostic methods are required for the assessment of such an important biomarker: urine albumin, creatinine, and their ratio (ACR) measured with quantitative assays are considered the gold standard. Routine dipstick methods (more rapid and at a lower cost) are intended for wide population screening. The aim of our study was to verify the reliability of an automated urinalysis dipstick method by comparing the results with the quantitative test of creatinine and albumin performed on a clinical chemistry platform. The first-morning voids of 249 patients who arrived from different departments were analyzed in the Central Laboratory of the University Hospital Policlinico Umberto I in Rome. We found a good correlation between the two assays, even though we observed that the dipstick assessment tends to overestimate the ACR's value, disclosing a higher number of false positives if compared to the reference method. As an important novelty in this study, we analyzed our data considering age (starting from pediatric to geriatric patients) and sex as variables for a sub-stratification of the participants. Our results show that positive values need to be confirmed with quantitative methods, especially in women and younger people, and that from samples that resulted as diluted at the dipstick assay, the ACR's values can be obtained if they are reanalyzed with quantitative assays. Moreover, patients with microalbuminuria (ACR 30-300 mg/g) or severe albumin urinary excretion (ACR > 300 mg/g) should be reanalyzed using quantitative methods to obtain a more reliable calculation of the ACR.
更多
查看译文
关键词
proteinuria,urine,reliable screening tool,automated platforms
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要