Simple and effective immobilization for radiation treatment of choroidal melanoma.

Medical dosimetry : official journal of the American Association of Medical Dosimetrists(2023)

引用 0|浏览6
暂无评分
摘要
At our institution, patients diagnosed with choroidal melanoma requiring external beam radiation therapy are treated with two 6 MV volumetric-modulated arcs delivering 50 Gy over 5 daily fractions. The patient is immobilized using an Orfit head and neck mask and is directed to look at a light emitting diode (LED) during CT simulation and treatment to minimize eye movement. Patient positioning is checked with cone beam computed tomography (CBCT) daily. Translational and rotational displacements greater than 1 mm or 1° off the planned isocenter position are corrected using a Hexapod couch. The aim of this study is to verify that the mask system provides adequate immobilization and to verify our 2-mm planning target volume (PTV) margins are sufficient. Residual displacements provided by pretreatment verification and post-treatment CBCT data sets were used to assess the impact of patient mobility during treatment on the reconstructed delivered dose to the target and organs at risk. The PTV margin calculated using van Herk's method was used to assess patient motion plus other factors that affect treatment position, such as kV-MV isocenter coincidence. Patient position variations were small and were shown to not cause significant dose variations between the planned and reconstructed dose to the target and organs at risk. The PTV margin analysis showed patient translational motion alone required a PTV margin of 1 mm. Given other factors that affect treatment delivery accuracy, a 2-mm PTV margin was shown to be sufficient for treatment of 95% of our patients with 100% of dose delivered to the GTV. The mask immobilization with LED focus is robust and we showed a 2-mm PTV margin is adequate with this technique.
更多
查看译文
关键词
Patient motion,Dose reconstruction,PTV margin assessment
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要