The Inactivation of SARS-CoV-2 in Critical Hospital Settings by Means of Ultraviolet C Lamps: Examples of Their Use and Some Practical Advice

Research Square (Research Square)(2021)

引用 0|浏览1
暂无评分
摘要
Abstract Background UltraViolet-C (UV-C) lamps may be used to supplement current hospital cleaning and disinfection of surfaces contaminated by SARS-CoV-2. Our aim is to provide some practical indications for the correct use of UV-C lamps. Methods We studied three UV-C lamps, measuring their spatial irradiance and emission over time. We quantify the error that is committed by calculating the irradiation time based exclusively on the technical data of the lamps or by making direct irradiance measurements. Finally, we tested specific dosimeters for UV-C. Results Our results show that the spatial emission of UV-C lamps is strongly dependent on the power of the lamps and on the design of their reflectors. Only by optimizing the positioning and calculating the exposure time correctly, is it possible to dispense the dose necessary to obtain SARS-CoV-2 inactivation. In the absence of suitable equipment for measuring irradiance, the calculated irradiation time can be underestimated. We therefore consider it precautionary to increase the calculated times by at least 20%. Conclusion To use UV-C lamps effectively, it is necessary to follow a few simple precepts when choosing, positioning and verifying the lamps. In the absence of instruments dedicated to direct verification of irradiance, photochromic UV-C dosimeters may represent a useful tool for easily verifying that a proper UV-C dose has been delivered.
更多
查看译文
关键词
critical hospital settings,sars-cov
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要