A new collimated multichannel modular detection system based on Silicon Drift Detectors

D. Cirrincione,M. Antonelli,G. Aquilanti,P. Bellutti,G. Bertuccio,G. Borghi,G. Cautero, F. Ficorella,M. Gandola, D. Giuressi,F. Mele,R.H. Menk,L. Olivi, G. Orzan, G. Pepponi, A. Picciotto, I. Rashevskaya, A. Rachevski, L. Stebel, G. Zampa

Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment(2023)

引用 0|浏览22
暂无评分
摘要
After the manufacture and delivery of a state-of-the-art detection system for the XRF-XAFS beamline of the synchrotron light source SESAME, a new and improved detection system was realized. This new multichannel modular detection system based on Silicon Drift Detectors consists of 8 monolithic multipixel arrays, each comprising 8 SDD cells with a total area of 570 mm2. As the previous one, this 64 channels integrated detection system includes ultra-low-noise front-end electronics, dedicated acquisition system, digital filtering, temperature control and stabilization. With respect to the SESAME version, the new instrument implements a tungsten collimation system yielding a total collimated sensitive area of 500 mm2. Optimized to work in an energy range of 3–30 keV, the system shows an overall energy resolution (sum of its 64 cells) below 180 eV FWHM at the 5.9 Mn Kα line at room temperature. We highlight the system performance and in particular the peak to background ratio, before and after the collimation of the sensors.
更多
查看译文
关键词
X-ray spectroscopy, Silicon Drift Detectors, XRF, XAFS, Detection system, Collimation